Reproducibility in light microscopy: Maintenance, standards and SOPs

Int J Biochem Cell Biol. 2017 Aug:89:120-124. doi: 10.1016/j.biocel.2017.06.008. Epub 2017 Jun 9.

Abstract

Light microscopy has grown to be a valuable asset in both the physical and life sciences. It is a highly quantitative method available in individual research laboratories and often centralized in core facilities. However, although quantitative microscopy is becoming a customary tool in research, it is rarely standardized. To achieve accurate quantitative microscopy data and reproducible results, three levels of standardization must be considered: (1) aspects of the microscope, (2) the sample, and (3) the detector. The accuracy of the data is only as reliable as the imaging system itself, thereby imposing the need for routine standard performance testing. Depending on the task some maintenance procedures should be performed once a month, some before each imaging session, while others conducted annually. This text should be implemented as a resource for researchers to integrate with their own standard operating procedures to ensure the highest quality quantitative microscopy data.

Keywords: Calibration; Microscopy; Quantitative; Standardize.

Publication types

  • Review
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Light*
  • Microscopy / instrumentation
  • Microscopy / methods*
  • Microscopy / standards*
  • Quality Control
  • Reference Standards
  • Reproducibility of Results