Effect of off-resonance on T1 saturation recovery measurement in inhomogeneous fields

J Magn Reson. 2017 Aug:281:31-43. doi: 10.1016/j.jmr.2017.05.003. Epub 2017 May 10.

Abstract

Saturation-recovery measurements with Carr-Purcell-Meiboom-Gill sequences are commonly employed to measure the longitudinal relaxation time constant, T1, in grossly inhomogeneous fields. We show that in general the off-resonant effect generates unexpected extra signals in the T1 measurement. In the present study, we derive a modified T1 kernel that accounts for this off-resonance effect quantitatively. The new kernel has been tested with numerical simulations and experiments, and excellent agreement is found.

Keywords: Dynamic equilibrium; Off-resonance; Saturation recovery.