Virtual Monoenergetic Imaging and Iodine Perfusion Maps Improve Diagnostic Accuracy of Dual-Energy Computed Tomography Pulmonary Angiography With Suboptimal Contrast Attenuation

Invest Radiol. 2017 Nov;52(11):659-665. doi: 10.1097/RLI.0000000000000387.

Abstract

Objectives: The aim of this study was to investigate the impact of virtual monoenergetic imaging (VMI+) and dual-energy computed tomography perfusion maps (DECT-PMs) on reader confidence and diagnostic accuracy in dual-energy computed tomography pulmonary angiography (DE-CTPA) studies with suboptimal contrast attenuation, compared with standard linearly blended reconstruction series.

Materials and methods: Dual-energy computed tomography pulmonary angiography examinations with suboptimal contrast attenuation of 68 patients with suspected pulmonary embolism (PE) were included in this institutional review board-approved retrospective study. Virtual monoenergetic imaging series at 40 keV, DECT-PM, and linearly blended images (M_0.6, 60% 90-kV spectrum) were reconstructed. Contrast-to-noise ratio and signal-to-noise ratio within the pulmonary trunk were calculated. Four independent radiologists assessed the presence of PE and their diagnostic confidence using 3 DE-CTPA reconstruction protocols: protocol 1, M_0.6 images; protocol 2, M_0.6 series and DECT-PM; and protocol 3, M_0.6, DECT-PM, and VMI+ series. Receiver operating characteristic (ROC) analysis was performed.

Results: Fourteen patients showed central and 29 segmental PE. Greater contrast-to-noise ratio and signal-to-noise ratio values were measured in VMI+ series at 40 keV in comparison to M_0.6 images (P < 0.001). Diagnostic accuracy for segmental PE detection was as follows: protocol 1 (69.1%); protocol 2 (86.8%); and protocol 3 (92.6%). Protocol 3 resulted in a significantly greater area under the curve for diagnosing segmental PE (0.991, P ≤ 0.033), compared with protocol 1 and 2 (0.897 and 0.951, respectively), and provided the highest diagnostic confidence (P < 0.001).

Conclusions: A reconstruction protocol including 40-keV VMI+ series and DECT-PM improves reader confidence and diagnostic accuracy for segmental PE detection compared with standard M_0.6 images in DE-CTPA with suboptimal contrast attenuation.

MeSH terms

  • Computed Tomography Angiography / methods*
  • Female
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Iodine
  • Male
  • Middle Aged
  • Pulmonary Artery / diagnostic imaging*
  • Radiography, Dual-Energy Scanned Projection / methods*
  • Reproducibility of Results
  • Retrospective Studies
  • Signal-To-Noise Ratio

Substances

  • Iodine