Benchmarking for On-Scalp MEG Sensors

IEEE Trans Biomed Eng. 2017 Jun;64(6):1270-1276. doi: 10.1109/TBME.2016.2599177.

Abstract

Objective: We present a benchmarking protocol for quantitatively comparing emerging on-scalp magnetoencephalography (MEG) sensor technologies to their counterparts in state-of-the-art MEG systems.

Methods: As a means of validation, we compare a high-critical-temperature superconducting quantum interference device (high Tc SQUID) with the low- Tc SQUIDs of an Elekta Neuromag TRIUX system in MEG recordings of auditory and somatosensory evoked fields (SEFs) on one human subject.

Results: We measure the expected signal gain for the auditory-evoked fields (deeper sources) and notice some unfamiliar features in the on-scalp sensor-based recordings of SEFs (shallower sources).

Conclusion: The experimental results serve as a proof of principle for the benchmarking protocol. This approach is straightforward, general to various on-scalp MEG sensors, and convenient to use on human subjects. The unexpected features in the SEFs suggest on-scalp MEG sensors may reveal information about neuromagnetic sources that is otherwise difficult to extract from state-of-the-art MEG recordings.

Significance: As the first systematically established on-scalp MEG benchmarking protocol, magnetic sensor developers can employ this method to prove the utility of their technology in MEG recordings. Further exploration of the SEFs with on-scalp MEG sensors may reveal unique information about their sources.

Publication types

  • Evaluation Study
  • Validation Study

MeSH terms

  • Benchmarking / standards*
  • Brain / physiopathology*
  • Electrodes / standards*
  • Equipment Design / standards
  • Equipment Failure Analysis / standards
  • Humans
  • Magnetoencephalography / instrumentation*
  • Magnetoencephalography / methods
  • Magnetoencephalography / standards*
  • Reference Values
  • Reproducibility of Results
  • Scalp / physiology*
  • Sensitivity and Specificity