Robust Wireless Power Transmission to mm-Sized Free-Floating Distributed Implants

IEEE Trans Biomed Circuits Syst. 2017 Jun;11(3):692-702. doi: 10.1109/TBCAS.2017.2663358. Epub 2017 May 19.

Abstract

This paper presents an inductive link for wireless power transmission (WPT) to mm-sized free-floating implants (FFIs) distributed in a large three-dimensional space in the neural tissue that is insensitive to the exact location of the receiver (Rx). The proposed structure utilizes a high-Q resonator on the target wirelessly powered plane that encompasses randomly positioned multiple FFIs, all powered by a large external transmitter (Tx). Based on resonant WPT fundamentals, we have devised a detailed method for optimization of the FFIs and explored design strategies and safety concerns, such as coil segmentation and specific absorption rate limits using realistic finite element simulation models in HFSS including head tissue layers, respectively. We have built several FFI prototypes to conduct accurate measurements and to characterize the performance of the proposed WPT method. Measurement results on 1-mm receivers operating at 60 MHz show power transfer efficiency and power delivered to the load at 2.4% and 1.3 mW, respectively, within 14-18 mm of Tx-Rx separation and 7 cm2 of brain surface.

MeSH terms

  • Electric Power Supplies*
  • Equipment Design
  • Finite Element Analysis
  • Humans
  • Prostheses and Implants*
  • Wireless Technology*