Development of a UHPLC-MS/MS (SRM) method for the quantitation of endogenous glucagon and dosed GLP-1 from human plasma

Bioanalysis. 2017 May;9(9):733-751. doi: 10.4155/bio-2017-0021. Epub 2017 May 10.

Abstract

Aim: The performance of glucagon and GLP-1 immunoassays is often poor, but few sensitive LC-MS/MS methods exist as alternatives.

Experimental: A multiplexed LC-MS/MS method using a 2D extraction technique was developed.

Results: The method was established for the quantitation of endogenous glucagon (LLOQ: 15 pg/ml) and dosed GLP-1 (LLOQ: 25 pg/ml) in human plasma, and is the first such method avoiding immunoenrichment. Specificity of endogenous glucagon quantitation was assured using a novel approach with a supercharging mobile phase additive to access a sensitive qualifier SRM. Endogenous glucagon concentrations were within the expected range, and showed good reproducibility after extended sample storage. A cross-validation against established immunoassays using physiological study samples demonstrated some similarities between methods.

Conclusion: The LC-MS/MS method offers a viable alternative to immunoassays for quantitation of endogenous glucagon, dosed glucagon and/or dosed GLP-1.

Keywords: GLP-1; LC–MS/MS; cross-validation; endogenous; glucagon; immunoassay; m-NBA; plasma; supercharging mobile phase additive.

Publication types

  • Validation Study

MeSH terms

  • Chromatography, High Pressure Liquid / methods*
  • Glucagon / blood*
  • Glucagon-Like Peptide 1 / blood*
  • Humans
  • Incretins / blood*
  • Limit of Detection
  • Tandem Mass Spectrometry / methods*

Substances

  • Incretins
  • Glucagon-Like Peptide 1
  • Glucagon