Immuno-PCR with digital readout

Biochem Biophys Res Commun. 2017 Jun 24;488(2):311-315. doi: 10.1016/j.bbrc.2017.04.162. Epub 2017 May 5.

Abstract

Immuno-PCR (IPCR) combines the versatile ELISA antigen detection with ultrasensitive PCR signal amplification, thereby enabling the highly sensitive detection of a broad range of targets with a typically very large dynamic detection range. The quantification of the antigen is usually achieved by real-time PCR, which provides a correlation between the target concentration and amplified DNA marker. We here report on the implementation of digital droplet PCR as a means for direct quantification of DNA copies to enable the highly sensitive detection of protein biomarkers. To this end, two alternative approaches, based on either magnetic microbead-based IPCR or a microplate-release IPCR were tested. The latter format worked well and revealed an extraordinary high robustness and sensitivity. While rtIPCR already fulfills typical immunoassay acceptance criteria, ddIPCR enables improved accuracy and precision of the assay because signal response and analyte concentrations are directly correlated. The utility of the novel ddIPCR technology is demonstrated at the example of two cytokines, interleukin 2 and interleukin 6 (IL2, IL6, respectively), with an overall average CV% of 5.0 (IL2) and 7.4 (IL6).

Keywords: Antibody; Biomarker; Digital droplet polymerase chain reaction; Immuno PCR; Immunoassays; Ultra-sensitive.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Antigens / analysis*
  • Cytokines / analysis*
  • Immunoassay / instrumentation
  • Immunoassay / methods*
  • Particle Size
  • Polymerase Chain Reaction / instrumentation
  • Polymerase Chain Reaction / methods*

Substances

  • Antigens
  • Cytokines