Detachable glass microelectrodes for recording action potentials in active moving organs

Am J Physiol Heart Circ Physiol. 2017 Jun 1;312(6):H1248-H1259. doi: 10.1152/ajpheart.00741.2016. Epub 2017 May 5.

Abstract

Here, we describe new detachable floating glass micropipette electrode devices that provide targeted action potential recordings in active moving organs without requiring constant mechanical constraint or pharmacological inhibition of tissue motion. The technology is based on the concept of a glass micropipette electrode that is held firmly during cell targeting and intracellular insertion, after which a 100-µg glass microelectrode, a "microdevice," is gently released to remain within the moving organ. The microdevices provide long-term recordings of action potentials, even during millimeter-scale movement of tissue in which the device is embedded. We demonstrate two different glass micropipette electrode holding and detachment designs appropriate for the heart (sharp glass microdevices for cardiac myocytes in rats, guinea pigs, and humans) and the brain (patch glass microdevices for neurons in rats). We explain how microdevices enable measurements of multiple cells within a moving organ that are typically difficult with other technologies. Using sharp microdevices, action potential duration was monitored continuously for 15 min in unconstrained perfused hearts during global ischemia-reperfusion, providing beat-to-beat measurements of changes in action potential duration. Action potentials from neurons in the hippocampus of anesthetized rats were measured with patch microdevices, which provided stable base potentials during long-term recordings. Our results demonstrate that detachable microdevices are an elegant and robust tool to record electrical activity with high temporal resolution and cellular level localization without disturbing the physiological working conditions of the organ.NEW & NOTEWORTHY Cellular action potential measurements within tissue using glass micropipette electrodes usually require tissue immobilization, potentially influencing the physiological relevance of the measurement. Here, we addressed this limitation with novel 100-µg detachable glass microelectrodes that can be precisely positioned to provide long-term measurements of action potential duration during unconstrained tissue movement.

Keywords: action potential duration; electrophysiology; glass micropipette electrodes; transmembrane potential.

Publication types

  • Video-Audio Media

MeSH terms

  • Action Potentials*
  • Animals
  • Equipment Design
  • Guinea Pigs
  • Humans
  • Microelectrodes*
  • Miniaturization
  • Movement*
  • Myocytes, Cardiac / physiology*
  • Neurons / physiology*
  • Patch-Clamp Techniques / instrumentation*
  • Rats, Sprague-Dawley
  • Time Factors