Ionised concentrations in calcium and magnesium buffers: Standards and precise measurement are mandatory

Prog Biophys Mol Biol. 2017 Jul:126:48-64. doi: 10.1016/j.pbiomolbio.2017.03.001. Epub 2017 Apr 25.

Abstract

In Ca2+ and Mg2+ buffer solutions the ionised concentrations ([X2+]) are either calculated or measured. Calculated values vary by up to a factor of seven due to the following four problems: The calculated [X2+] in buffers are so inconsistent that calculation is not an option. Until standards are available, the [X2+] in the buffers must be measured. The Ligand Optimisation Method is an accurate and independently verified method of doing this (McGuigan & Stumpff, Anal. Biochem. 436, 29, 2013). Lack of standards means it is not possible to compare the published [Ca2+] in the nmolar range, and the apparent constant (K/) values for Ca2+ and Mg2+ binding to intracellular ligands amongst different laboratories. Standardisation of Ca2+/Mg2+ buffers is now essential. The parameters to achieve this are proposed.

Keywords: Ca(2+)/Mg(2+) buffers; Calculated ionised concentrations; Ionised concentrations; Ligand optimisation method; Measured ionised concentrations; Standard Ca(2+)/Mg(2+) buffers.

Publication types

  • Published Erratum