External closed-circuit cooling system for management of patients after device implantation: A feasibility study

Int J Cardiol. 2017 Aug 15:241:235-237. doi: 10.1016/j.ijcard.2017.04.048. Epub 2017 Apr 19.

Abstract

Background: In the first 24h after pacemaker or implantable cardioverter/defibrillator (ICD) implantation or replacement, the occurrence of hematoma and pain in the surgically treated region is not infrequent and may result in re-intervention and/or more severe complications, such as infections. Currently, the post-implant phase management is very empiric. The aim of this study was to test the clinical applicability and usefulness of an external close-circuit cooling system for the management of the early post-implant period in patients with high risk of hematoma due to anticoagulant and/or antiplatelet therapy.

Methods: We studied 135 patients (78M; 71±11years) with high risk of hematoma occurrence after pace-maker (63 patients) or ICD (72 patients) implantation or replacement. Immediately after the intervention, a closed-circuit cooling system (CAREPACE™ system, Zamar, Italy) was externally applied on the pre-pectoral region to each patient and maintained for 24h. The system has a compressive pad and a refrigerating circuit in which non-toxic glycolic fluid is pumped. The fluid temperature was set and kept at 5°C for the whole period.

Results: The compressive and cooling effect of the system was well tolerated by all the patients at the temperature set. Four patients complained of noise due to machine operation, but in none the treatment was interrupted. The average length of hospital stay was 2.8±0.4days. No clinically significant hematoma was observed at discharge and after one month follow-up visit.

Conclusions: This new system can be used for the management of the early phase after device implantation or replacement and appears clinically useful and well tolerated. Further studies on a larger scale are needed to test the potential reduction of post-intervention complications and the cost-effectiveness of this device.

Keywords: Haematoma; ICD; Pacing.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Cold Temperature*
  • Defibrillators, Implantable / adverse effects
  • Defibrillators, Implantable / standards*
  • Disease Management*
  • Feasibility Studies
  • Female
  • Follow-Up Studies
  • Hematoma / etiology
  • Hematoma / prevention & control
  • Humans
  • Male
  • Middle Aged
  • Pacemaker, Artificial / adverse effects
  • Pacemaker, Artificial / standards*
  • Risk Factors