Optical performance characterization of light-logging actigraphy dosimeters

J Opt Soc Am A Opt Image Sci Vis. 2017 Apr 1;34(4):545-557. doi: 10.1364/JOSAA.34.000545.

Abstract

There are several wearable products specially developed or marketed for studying sleep, circadian rhythms, and light levels. However, new recommendations relating to human physiological responses to light have changed what measurements researchers may demand. The performances of 11 light-logging dosimeters from eight manufacturers were compared. The directional and spectral sensitivities, linearity, dynamic range, and resolution were tested for seven models, and compared along with other published data. The sample mainly comprised light-logging actigraphy dosimeters wearable as badges, in accordance with measurement protocols for larger-scale field studies. A proposed standard for optical performance assessments is set out.

Publication types

  • Comparative Study

MeSH terms

  • Actigraphy / instrumentation*
  • Circadian Rhythm / physiology*
  • Humans
  • Optical Phenomena*
  • Radiation Dosimeters / standards*
  • Sleep / physiology*