Evaluation of a Compact Hybrid Brain-Computer Interface System

Biomed Res Int. 2017:2017:6820482. doi: 10.1155/2017/6820482. Epub 2017 Mar 8.

Abstract

We realized a compact hybrid brain-computer interface (BCI) system by integrating a portable near-infrared spectroscopy (NIRS) device with an economical electroencephalography (EEG) system. The NIRS array was located on the subjects' forehead, covering the prefrontal area. The EEG electrodes were distributed over the frontal, motor/temporal, and parietal areas. The experimental paradigm involved a Stroop word-picture matching test in combination with mental arithmetic (MA) and baseline (BL) tasks, in which the subjects were asked to perform either MA or BL in response to congruent or incongruent conditions, respectively. We compared the classification accuracies of each of the modalities (NIRS or EEG) with that of the hybrid system. We showed that the hybrid system outperforms the unimodal EEG and NIRS systems by 6.2% and 2.5%, respectively. Since the proposed hybrid system is based on portable platforms, it is not confined to a laboratory environment and has the potential to be used in real-life situations, such as in neurorehabilitation.

MeSH terms

  • Adult
  • Brain-Computer Interfaces*
  • Electroencephalography
  • Female
  • Humans
  • Male
  • Mathematics
  • Prefrontal Cortex / physiology*
  • Psychomotor Performance / physiology
  • Signal Processing, Computer-Assisted
  • Spectroscopy, Near-Infrared / instrumentation