Graphene-Enabled Electrodes for Electrocardiogram Monitoring

Nanomaterials (Basel). 2016 Aug 23;6(9):156. doi: 10.3390/nano6090156.

Abstract

The unique parameters of graphene (GN)-notably its considerable electron mobility, high surface area, and electrical conductivity-are bringing extensive attention into the wearable technologies. This work presents a novel graphene-based electrode for acquisition of electrocardiogram (ECG). The proposed electrode was fabricated by coating GN on top of a metallic layer of a Ag/AgCl electrode using a chemical vapour deposition (CVD) technique. To investigate the performance of the fabricated GN-based electrode, two types of electrodes were fabricated with different sizes to conduct the signal qualities and the skin-electrode contact impedance measurements. Performances of the GN-enabled electrodes were compared to the conventional Ag/AgCl electrodes in terms of ECG signal quality, skin-electrode contact impedance, signal-to-noise ratio (SNR), and response time. Experimental results showed the proposed GN-based electrodes produced better ECG signals, higher SNR (improved by 8%), and lower contact impedance (improved by 78%) values than conventional ECG electrodes.

Keywords: Raman spectroscopy; contact impedance; dry electrodes; electrocardiogram (ECG); graphene; long-term monitoring.