Assessment of Reproducibility of Laser Electrospray Mass Spectrometry using Electrospray Deposition of Analyte

J Am Soc Mass Spectrom. 2017 May;28(5):880-886. doi: 10.1007/s13361-017-1622-2. Epub 2017 Mar 15.

Abstract

A nonresonant, femtosecond (fs) laser is employed to desorb samples of Victoria blue deposited on stainless steel or indium tin oxide (ITO) slides using either electrospray deposition (ESD) or dried droplet deposition. The use of ESD resulted in uniform films of Victoria blue whereas the dried droplet method resulted in the formation of a ring pattern of the dye. Laser electrospray mass spectrometry (LEMS) measurements of the ESD-prepared films on either substrate were similar and revealed lower average relative standard deviations for measurements within-film (20.9%) and between-films (8.7%) in comparison to dried droplet (75.5% and 40.2%, respectively). The mass spectral response for ESD samples on both substrates was linear (R2 > 0.99), enabling quantitative measurements over the selected range of 7.0 × 10-11 to 2.8 × 10-9 mol, as opposed to the dried droplet samples where quantitation was not possible (R2 = 0.56). The limit of detection was measured to be 210 fmol. Graphical Abstract ᅟ.

Keywords: Electrospray deposition; Femtosecond laser vaporization; Quantitative analysis.