Effect of vibration frequency on biopsy needle insertion force

Med Eng Phys. 2017 May:43:71-76. doi: 10.1016/j.medengphy.2017.02.011. Epub 2017 Mar 1.

Abstract

Needle insertion is critical in many clinical medicine procedures, such as biopsy, brachytherapy, and injection therapy. A platform with two degrees of freedom was set up to study the effect of vibration frequency on needle insertion force. The gel phantom deformation at the needle cutting edge and the Voigt model are utilized to develop a dynamic model to explain the relationship between the insertion force and needle-tip velocity. The accuracy of this model was verified by performing needle insertions into phantom gel. The effect of vibration on insertion force can be explained as the vibration increasing the needle-tip velocity and subsequently increasing the insertion force. In a series of needle insertion experiments with different vibration frequencies, the peak forces were selected for comparison to explore the effect of vibration frequency on needle insertion force. The experimental results indicate that the insertion force at 500Hz increases up to 17.9% compared with the force at 50Hz.

Keywords: Insertion force; Needle insertion; Vibration; Voigt model.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biopsy, Needle / instrumentation*
  • Mechanical Phenomena*
  • Vibration*