Light curing procedures - performance, knowledge level and safety awareness among dentists

J Dent. 2017 Mar:58:67-73. doi: 10.1016/j.jdent.2017.02.002. Epub 2017 Feb 6.

Abstract

Objectives: This study aimed to investigate dentists' exposure to curing light and to obtain information about the dentists' knowledge on practical use and technical features of their curing lights as well as their safety awareness.

Methods: A pre-coded questionnaire was sent electronically to all dentists (n=1313) in the Public Dental Service (PDS) in Norway in 2015.

Results: The Response rate was 55.8%. The dentists spent on average 57.5% of their working days placing restorations, ranging from 1 to 30 (mean 7.7, SD 3.6) restorations per day. The average length of light curing one normal layer of composite was 27s. The longest individual mean curing time per day was about 100 times higher than that of the lowest. The mean curing time for lamps of the lower reported irradiances was similar to the time representing exceedance of international guidelines for limit values for blue light to the eyes. Almost one-third of the dentists used inadequate eye protection against blue light. The odds of using adequate eye protection were significantly higher among young dentists (p<0.01). The majority of the respondents (78.3%) were unaware of the irradiance value of their curing lights, thus rendering the curing time uncertain. More dentists in this group did not perform regular maintenance of their curing lights compared with all respondents (17.1% vs. 3.3%, p<0.01).

Conclusions: This study revealed considerable variations among Norwegian dentists in the Public Dental Service with respect to performance of light curing of restorations, safety awareness and technical knowledge of the curing light.

Clinical significance: The questionnaire study identifies specific knowledge gaps among Norwegian dentists with regard to curing lights and use of personal protection. Today's dependence on technology in dentistry necessitates that the operator possesses knowledge of essential technical specifications and safe use of devices and instruments routinely used in dental treatment.

Keywords: Dental curing lights; Dental restoration; Maximum permissible exposure time; Resin composite; Visible light.

MeSH terms

  • Adult
  • Aged
  • Composite Resins / chemistry
  • Composite Resins / radiation effects
  • Curing Lights, Dental / adverse effects
  • Curing Lights, Dental / standards
  • Curing Lights, Dental / statistics & numerical data*
  • Dental Offices
  • Dental Restoration, Permanent / statistics & numerical data
  • Dentists / education*
  • Dentists / psychology*
  • Eye / radiation effects
  • Female
  • Health Knowledge, Attitudes, Practice
  • Humans
  • Light-Curing of Dental Adhesives
  • Maintenance
  • Male
  • Maximum Allowable Concentration
  • Middle Aged
  • Norway
  • Radiation Dosage
  • Radiation Protection / standards
  • Risk Assessment
  • Safety
  • Surveys and Questionnaires
  • Time Factors

Substances

  • Composite Resins