A Comparison of Three Quantitative Methods to Estimate G6PD Activity in the Chittagong Hill Tracts, Bangladesh

PLoS One. 2017 Jan 25;12(1):e0169930. doi: 10.1371/journal.pone.0169930. eCollection 2017.

Abstract

Background: Glucose-6-phosphate-dehydrogenase-deficiency (G6PDd) is a major risk factor for primaquine-induced haemolysis. There is a need for improved point-of-care and laboratory-based G6PD diagnostics to unsure safe use of primaquine.

Methods: G6PD activities of participants in a cross-sectional survey in Bangladesh were assessed using two novel quantitative assays, the modified WST-8 test and the CareStart™ G6PD Biosensor (Access Bio), The results were compared with a gold standard UV spectrophotometry assay (Randox). The handheld CareStart™ Hb instrument (Access Bio) is designed to be a companion instrument to the CareStart™ G6PD biosensor, and its performance was compared to the well-validated HemoCue™ method. All quantitative G6PD results were normalized with the HemoCue™ result.

Results: A total of 1002 individuals were enrolled. The adjusted male median (AMM) derived by spectrophotometry was 7.03 U/g Hb (interquartile range (IQR): 5.38-8.69), by WST-8 was 7.03 U/g Hb (IQR: 5.22-8.16) and by Biosensor was 8.61 U/g Hb (IQR: 6.71-10.08). The AMM between spectrophotometry and WST-8 did not differ (p = 1.0) but differed significantly between spectrophotometry and Biosensor (p<0.01). Both, WST-8 and Biosensor were correlated with spectrophotometry (rs = 0.5 and rs = 0.4, both p<0.001). The mean difference in G6PD activity was -0.12 U/g Hb (95% limit of agreement (95% LoA): -5.45 to 5.20) between spectrophotometry and WST-8 and -1.74U/g Hb (95% LoA: -7.63 to 4.23) between spectrophotometry and Biosensor. The WST-8 identified 55.1% (49/89) and the Biosensor 19.1% (17/89) of individuals with G6PD activity <30% by spectrophotometry. Areas under the ROC curve did not differ significantly for the WST-8 and Biosensor irrespective of the cut-off activity applied (all p>0.05). Sensitivity and specificity for detecting G6PD activity <30% was 0.55 (95% confidence interval (95%CI): 0.44-0.66) and 0.98 (95%CI: 0.97-0.99) respectively for the WST-8 and 0.19 (95%CI: 0.12-0.29) and 0.99 (95%CI: 0.98-0.99) respectively for the Biosensor. Hb concentrations measured by HemoCue™ and CareStart™ Hb were strongly correlated (rs = 0.8, p<0.001, mean difference = 0.09 g Hb/dL, 95% LoA: -2.15 to 2.34).

Conclusion: WST-8 and the CareStart™ G6PD Biosensor represent advances in G6PD diagnostics in resource poor settings, but will require further development before clinical deployment. The CareStart™ Hb instrument produced a precise measure of haemoglobin concentration.

Publication types

  • Comparative Study
  • Validation Study

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Bangladesh / epidemiology
  • Biosensing Techniques* / instrumentation
  • Child
  • Child, Preschool
  • Colorimetry* / instrumentation
  • Cross-Sectional Studies
  • Endemic Diseases
  • Female
  • Glucosephosphate Dehydrogenase / blood*
  • Glucosephosphate Dehydrogenase Deficiency / blood
  • Glucosephosphate Dehydrogenase Deficiency / epidemiology*
  • Glucosephosphate Dehydrogenase Deficiency / ethnology
  • Hemoglobins / analysis
  • Humans
  • Malaria / epidemiology
  • Male
  • Middle Aged
  • Point-of-Care Systems
  • Predictive Value of Tests
  • ROC Curve
  • Sensitivity and Specificity
  • Spectrophotometry, Ultraviolet* / instrumentation
  • Young Adult

Substances

  • Hemoglobins
  • Glucosephosphate Dehydrogenase