Poor Validity of Noninvasive Hemoglobin Measurements by Pulse Oximetry Compared with Conventional Absorptiometry in Children in Côte d'Ivoire

Am J Trop Med Hyg. 2017 Jan 11;96(1):217-220. doi: 10.4269/ajtmh.16-0505. Epub 2016 Nov 14.

Abstract

Anemia remains a major public health issue in many African communities. We compared a novel, commercially available noninvasive hemoglobin (Hb)-measuring device to direct Hb measurements by finger-prick samples in a pediatric cohort in rural Côte d'Ivoire. Noninvasive Hb measurements were attempted in 191 children 2-15 years of age and obtained in 102 (53.5%) children. The median Hb for the 102 children was 12.0 g/dL (interquartile range [IQR] = 11.3-12.7 g/dL) for conventional absorptiometry and 13.3 g/dL (IQR = 12.1-14.2 g/dL) for noninvasive measurements. A Bland-Altman analysis demonstrated a median bias of +1.1 g/dL (IQR = 0.4-2.0 g/dL), with greater overestimation of Hb by noninvasive testing occurring at low Hb values. This overestimation of the noninvasive Hb-measuring device to direct Hb measurements persisted across preschool- and school-aged children, and both sexes. The Pearson correlation coefficient was 0.50 for children 4-9 years of age, and 0.33 for children 10-15 years of age. Further study and development of noninvasive Hb devices is necessary prior to implementation in African pediatric populations.

MeSH terms

  • Adolescent
  • Anemia / diagnosis*
  • Anemia / epidemiology
  • Child
  • Child, Preschool
  • Cote d'Ivoire / epidemiology
  • Female
  • Hemoglobins*
  • Humans
  • Male
  • Oximetry / methods*
  • Reproducibility of Results