Diagnostic performance of a high-sensitive troponin T assay and a troponin T point of care assay in the clinical routine of an Emergency Department: A clinical cohort study

Int J Cardiol. 2017 Mar 1:230:454-460. doi: 10.1016/j.ijcard.2016.12.085. Epub 2016 Dec 21.

Abstract

Background: A point of care test (POCT) for troponin T (TnT) in the Emergency Department (ED) was compared to a high-sensitivity TnT (hsTnT) central laboratory test (CLT) to determine the influence of test system and different cut-off values on the diagnostic performance in patients with suspected acute coronary syndrome (ACS) under routine conditions.

Methods: All patients with routine TnT testing in the ED were enrolled. Only internal medicine patients without STEMI and with both troponin values were analyzed. TnT was measured with a contemporary sensitive POCT assay in the ED and with a hs-assay in the central laboratory. The diagnostic performance was analyzed at two different cut-off points (99th percentile and conventional rule-in cut-offs). Primary endpoint was the diagnosis of NSTEMI.

Results: Of all patients (n=3423), 3.6% had a diagnosis of NSTEMI (n=124). For the hsTnT assay, 28.4% of all values were at or below the lower limit of detection (LOD) as compared to 75.7% of the POC-TnT-values. The area under the receiver operating curves did not differ significantly between the assays (hsTnT: 0.912(95%-CI: 0.884-0.940); POC-TnT: 0.896(95%-CI: 0.859-0.933)). The diagnostic performance was very similar for both assays: the positive predictive value was below 50% for troponin values below 100ng/L and hardly increased for values between 100 and 600ng/L for hs and conventional assays.

Conclusions: In our cohort of emergency patients, the diagnostic performance of conventional POC-testing was comparable to hsTnT. A 99th percentile cut-off may be useful for rule-out of NSTEMI, but seems limited for routine rule-in strategies.

Keywords: Acute coronary syndrome; Diagnostic performance; High sensitive troponin; Non-ST elevation myocardial infarction; Point of care testing.

MeSH terms

  • Acute Coronary Syndrome / blood
  • Acute Coronary Syndrome / complications
  • Acute Coronary Syndrome / diagnosis*
  • Adult
  • Aged
  • Cohort Studies
  • Emergency Service, Hospital*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Non-ST Elevated Myocardial Infarction / blood*
  • Non-ST Elevated Myocardial Infarction / diagnosis*
  • Non-ST Elevated Myocardial Infarction / etiology
  • Point-of-Care Systems*
  • Sensitivity and Specificity
  • Troponin T / blood*

Substances

  • Troponin T