A Simple Interfacial Platform for Homogeneous Electrochemical Immunoassays Using a Poly(Vinylimidazole)-Modified Electrode

Sensors (Basel). 2016 Dec 29;17(1):54. doi: 10.3390/s17010054.

Abstract

In this study, a homogeneous method featuring simple, one-step detection was developed to analyze hippuric acid (HA), a major metabolite of toluene. High sensitivity was achieved with the facile immobilization of poly(vinylimidazole) (PVI) on an indium tin oxide (ITO) electrode. Using a previously developed approach, pentacyanoferrate was coordinated with pyridyl-N ligands, and the redox-active Fe(II/III) centers were bound to Ni(II) ions on the electrode via electrostatic cyanide bridges. The detection was accomplished by the competitive binding of free HA and pentacyanoferrate-(4-aminomethylpyridine-hippuric acid) (Fe-HA, the electron transfer mediator) to the HA antibody on the Ni(II) ions-modified PVI-ITO (Ni-PVI-ITO) electrode. The electrical and physicochemical characterization of the electrode was carried out by cyclic voltammetry, differential pulse voltammetry, field emission scanning electron microscopy, and X-ray photoelectron spectroscopy. At low mediator concentrations, the electrical signals were proportional to the HA concentration between 0.1 µg/mL and 1.0 mg/mL. The same method may be extended to other small organic molecules.

Keywords: electrochemical biosensor; hippuric acid; immunoassay; mediator; nickel ion.

MeSH terms

  • Biosensing Techniques / methods*
  • Electrochemical Techniques / methods
  • Electrodes
  • Hippurates / chemistry
  • Immunoassay / methods
  • Oxidation-Reduction
  • Photoelectron Spectroscopy / methods

Substances

  • Hippurates
  • hippuric acid