Single cardiac cycle three-dimensional intracoronary optical coherence tomography

Biomed Opt Express. 2016 Nov 1;7(12):4847-4858. doi: 10.1364/BOE.7.004847. eCollection 2016 Dec 1.

Abstract

While high-speed intracoronary optical coherence tomography (OCT) provides three-dimensional (3D) visualization of coronary arteries in vivo, imaging speeds remain insufficient to avoid motion artifacts induced by heartbeat, limiting the clinical utility of OCT. In this paper, we demonstrate development of a high-speed intracoronary OCT system (frame rate: 500 frames/s, pullback speed: 100 mm/s) along with prospective electrocardiogram (ECG) triggering technology, which enabled volumetric imaging of long coronary segments within a single cardiac cycle (70 mm pullback in 0.7 s) with minimal cardiac motion artifact. This technology permitted detailed visualization of 3D architecture of the coronary arterial wall of a swine in vivo and fine structure of the implanted stent.

Keywords: (170.2150) Endoscopic imaging; (170.3880) Medical and biological imaging; (170.4500) Optical coherence tomography.