Options to Expand HIV Viral Load Testing in South Africa: Evaluation of the GeneXpert® HIV-1 Viral Load Assay

PLoS One. 2016 Dec 16;11(12):e0168244. doi: 10.1371/journal.pone.0168244. eCollection 2016.

Abstract

Background: Expansion of HIV viral load (VL) testing services are required to meet increased targets for monitoring patients on antiretroviral treatment. South Africa currently tests >4million VLs per annum in 16 highly centralised, automated high-throughput laboratories. The Xpert HIV-1 VL assay (Cepheid) was evaluated against in-country predicates, the Roche Cobas Taqmanv2 and Abbott HIV-1RT, to investigate options for expanding VL testing using GeneXpert's random access, polyvalent capabilities and already established footprint in South Africa with the Xpert MTB/RIF assay (207 sites). Additionally, the performance of Xpert HIV-1VL on alternative, off-label specimen types, Dried Blood Spots (DBS) and whole blood, was investigated.

Method: Precision, accuracy (agreement) and clinical misclassification (1000cp/ml) of Xpert HIV-1VL plasma was compared to Taqmanv2 (n = 155) and Abbott HIV-1 RT (n = 145). Misclassification of Xpert HIV-1VL was further tested on DBS (n = 145) and whole blood (n = 147).

Results: Xpert HIV-1VL demonstrated 100% concordance with predicate platforms on a standardised frozen, plasma panel (n = 42) and low overall percentage similarity CV of 1.5% and 0.9% compared to Taqmanv2 and Abbott HIV-1 RT, respectively. On paired plasma clinical specimens, Xpert HIV-1VL had low bias (SD 0.32-0.37logcp/ml) and 3% misclassification at the 1000cp/ml threshold compared to Taqmanv2 (fresh) and Abbott HIV-1 RT (frozen), respectively. Xpert HIV-1VL on whole blood and DBS increased misclassification (upward) by up to 14% with increased invalid rate. All specimen testing was easy to perform and compatible with concurrent Xpert MTB/RIF Tuberculosis testing on the same instrument.

Conclusion: The Xpert HIV-1VL on plasma can be used interchangeably with existing predicate platforms in South Africa. Whole blood and DBS testing requires further investigation, but polyvalency of the GeneXpert offers a solution to extending VL testing services.

Publication types

  • Evaluation Study

MeSH terms

  • Adult
  • Antiretroviral Therapy, Highly Active
  • Female
  • HIV Infections / diagnosis*
  • HIV Infections / drug therapy
  • HIV Infections / epidemiology
  • HIV-1
  • Humans
  • Male
  • Mass Screening / methods*
  • Middle Aged
  • Molecular Diagnostic Techniques / methods*
  • RNA, Viral / analysis
  • RNA, Viral / blood*
  • Sensitivity and Specificity
  • South Africa / epidemiology
  • Specimen Handling / methods
  • Viral Load / methods*

Substances

  • RNA, Viral

Grants and funding

This publication was made possible by the generous financial support of the Grand Challenges Canada POC implementation grant (grant 0007-02-01-01-01). The funder had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.