Rotating and translating anthropomorphic head voxel models to establish an horizontal Frankfort plane for dental CBCT Monte Carlo simulations: a dose comparison study

Phys Med Biol. 2016 Dec 21;61(24):N681-N696. doi: 10.1088/1361-6560/61/24/N681. Epub 2016 Nov 28.

Abstract

In order to carry out Monte Carlo (MC) dosimetry studies, voxel phantoms, modeling human anatomy, and organ-based segmentation of CT image data sets are applied to simulation frameworks. The resulting voxel phantoms preserve patient CT acquisition geometry; in the case of head voxel models built upon head CT images, the head support with which CT scanners are equipped introduces an inclination to the head, and hence to the head voxel model. In dental cone beam CT (CBCT) imaging, patients are always positioned in such a way that the Frankfort line is horizontal, implying that there is no head inclination. The orientation of the head is important, as it influences the distance of critical radiosensitive organs like the thyroid and the esophagus from the x-ray tube. This work aims to propose a procedure to adjust head voxel phantom orientation, and to investigate the impact of head inclination on organ doses in dental CBCT MC dosimetry studies. The female adult ICRP, and three in-house-built paediatric voxel phantoms were in this study. An EGSnrc MC framework was employed to simulate two commonly used protocols; a Morita Accuitomo 170 dental CBCT scanner (FOVs: 60 × 60 mm2 and 80 × 80 mm2, standard resolution), and a 3D Teeth protocol (FOV: 100 × 90 mm2) in a Planmeca Promax 3D MAX scanner. Result analysis revealed large absorbed organ dose differences in radiosensitive organs between the original and the geometrically corrected voxel models of this study, ranging from -45.6% to 39.3%. Therefore, accurate dental CBCT MC dose calculations require geometrical adjustments to be applied to head voxel models.

MeSH terms

  • Adult
  • Anthroposophy
  • Child
  • Child, Preschool
  • Cone-Beam Computed Tomography / methods*
  • Female
  • Head / diagnostic imaging*
  • Humans
  • Jaw / diagnostic imaging
  • Male
  • Models, Dental*
  • Monte Carlo Method*
  • Phantoms, Imaging*
  • Radiation Dosage
  • Tomography Scanners, X-Ray Computed
  • Tomography, X-Ray Computed