Prevalidation trial for a novel in vitro eye irritation test using the reconstructed human cornea-like epithelial model, MCTT HCE™

Toxicol In Vitro. 2017 Mar:39:58-67. doi: 10.1016/j.tiv.2016.11.010. Epub 2016 Nov 19.

Abstract

Here, we report the results of a prevalidation trial for an in vitro eye irritation test (EIT) using the reconstructed human cornea-like epithelium, MCTT HCE™. The optimal cutoff to determine irritation in the prediction model was established at 35% with the receiver operation characteristics(ROC) curve for 126 substances. Within-lab(WL) and between-lab(BL) reproducibility was tested for 20 reference substances by 3 participating laboratories. Viability data described by mean±SD or ±1/2 difference between duplicate wells, and scatter plots, demonstrated the WL/BL consistency. WL/BL concordance with the binary decision, whether non-irritant or irritant was estimated to be 85-95% and 95%, respectively. WL/BL reproducibility of viability data was further supported by a strong correlation(ICC, r>0.9). WL/BL agreement of binary decisions was also examined by Fleiss' Kappa statistics, which showed a strong level of agreement (>0.78), nevertheless weaker than the reproducibility of the viability. The EIT with MCTT HCE™ exhibited a sensitivity of 82.2% (60/73), a specificity of 81.1% (43/53), and an accuracy of 81.8% (103/126) for 126 reference substances (for liquids; a sensitivity of 100% (47/47), a specificity of 70.6% (24/34), and an accuracy of 87.7% (71/81), and for solids, a sensitivity of 50% (13/26), a specificity of 100% (19/19), and an accuracy of 71.1% (32/45), suggesting that the accuracy is satisfactory but the sensitivity needs improvement, which shall be addressed through correcting the poor sensitivity for solid substances in future full validation trials.

Keywords: Eye irritation test; MCTT HCE™; Predictive capacity; Reproducibility; Statistical analysis.

MeSH terms

  • Animal Testing Alternatives
  • Cell Survival / drug effects
  • Epithelium, Corneal / drug effects*
  • Humans
  • In Vitro Techniques
  • Irritants / toxicity*
  • Reproducibility of Results
  • Sensitivity and Specificity

Substances

  • Irritants