Zero field PDD and TMR data for unflattened beams in conventional linacs: A tool for independent dose calculations

Phys Med. 2016 Dec;32(12):1621-1627. doi: 10.1016/j.ejmp.2016.11.008. Epub 2016 Nov 18.

Abstract

Purpose: To investigate the applicability of the formalism described in BJR supplement n.25 for Flattening Filter Free (FFF) beams in determining the zero-field tissue maximum ratio (TMR) for an independent calculation method of Percentage Depth Doses (PDDs) and relative dose factors (RDFs) at different experimental setups.

Methods: Experimental PDDs for field size from 40×40cm2 to 2×2cm2 with Source Surface Distance (SSD) 100cm were acquired. The normalized peak scatter factor for each square field was obtained by fitting experimental RDFs in water and collimator factors (CFs) in air. Maximum log-likelihood methods were used to extract fit parameters in competing models and the Bayesian Information Criterion was used to select the best one. In different experimental setups additional RDFs and TPR1020s for field sizes other than reference field were measured and Monte Carlo simulations of PDDs at SSD 80cm were carried out to validate the results. PDD agreements were evaluated by gamma analysis.

Results: The BJR formalism allowed to predict the PDDs obtained with MC within 2%/2mm at SSD 80cm from 100% down to 50% of the maximum dose. The agreement between experimental TPR1020s and RDFs values at SSD=90cm and BJR calculations were within 1% for field sizes greater than 5×5cm2 while it was within 3% for fields down to 2×2cm2.

Conclusions: BJR formalism can be used for FFF beams to predict PDD and RDF at different SSDs and can be used for independent MU calculations.

Keywords: Analytical dose calculation; Flattening Filter Free; Small fields; TPR.