Feasibility of Fraction Collection in HPLC Systems with Evaporative Light Scattering Detector: Analysis of Pectinatella magnifica

Molecules. 2016 Nov 8;21(11):1495. doi: 10.3390/molecules21111495.

Abstract

The use of a liquid chromatography (LC) splitter inserted between an HPLC column and an evaporative light scattering detector (ELSD) is described. This paper aims to show the feasibility of using the splitter in an HPLC-ELSD system to fractionate a model mixture of analytes, namely salicin (2-(hydroxymethyl)-phenyl-β-d-glucopyranoside) and glucose. The retention factors and efficiency of the separation were studied under various temperatures and water contents in the mobile phase in order to clarify the mechanism of polyols separation on a diol column under the conditions of hydrophilic liquid chromatography (HILIC). Finally, the system was applied to a biological sample (a lyophilized colony gel of Pectinatella magnifica), where the presence of fructose and glucose was confirmed.

Keywords: Pectinatella magnifica; evaporative light scattering detector; fraction collection; hydrophilic interaction liquid chromatography; mixed-mode chromatography.

MeSH terms

  • Benzyl Alcohols / chemistry*
  • Chromatography, High Pressure Liquid / methods
  • Fructose / chemistry
  • Glucose / chemistry*
  • Glucosides / chemistry*
  • Hydrophobic and Hydrophilic Interactions
  • Scattering, Radiation
  • Temperature
  • Water / chemistry

Substances

  • Benzyl Alcohols
  • Glucosides
  • Water
  • Fructose
  • salicin
  • Glucose