The influence of sample freezing at - 80 °C for 2-12 weeks on glycated haemoglobin (HbA1c) concentration assayed by HPLC method on Bio-Rad D-10® auto analyzer

Biochem Med (Zagreb). 2016 Oct 15;26(3):346-352. doi: 10.11613/BM.2016.038.

Abstract

Introduction: The aim of the study was to evaluate the effect of a single freeze/thaw cycle on HbA1c concentrations measured by commercially available HPLC method.

Materials and methods: Study included 128 whole blood samples collected from diabetic patients (N = 60) and healthy volunteers (N = 68). HbA1c concentrations were measured in fresh blood samples. Then samples were frozen at - 80 °C for up to 12 weeks. HbA1c was assayed by ion-exchange HPLC method on Bio-Rad D-10® analyzer. Variables were compared using Wilcoxon and ANOVA Kruskal-Wallis tests. Bias between HbA1c measured in fresh and frozen samples was calculated. The comparability of HbA1c concentrations was assessed by Bland-Altman plot.

Results: Median (IQR) HbA1c concentration was 45.3 (36.6-61.2) mmol/mol for fresh and 45.3 (36.6-60.6) mmol/mol for frozen/thawed samples. No significant difference in HbA1c concentrations was found comparing fresh and frozen/thawed samples (P = 0.070) in the whole group, as well as in healthy and diabetic subjects. The median calculated bias between fresh and frozen/thawed samples was 0% in whole group and healthy subjects, and 1.19% in diabetic patients. No significant difference was found between the biases according to baseline HbA1c values (P = 0.150). The Bland-Altman plot analysis showed a positive bias of 0.4% (95% CI: - 2.8 - 3.7%), which indicates high compliance between HbA1c values and no relevant influence of sample freezing on clinical significance of HbA1c measurement.

Conclusions: Storage for up to 12 weeks at - 80 °C with a single freeze/thaw cycle does not affect HbA1c concentrations measured with HPLC method on Bio-Rad D-10® analyzer.

Keywords: diabetes; glycated haemoglobin; high performance liquid chromatography; preanalytical phase; sample storage.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Case-Control Studies
  • Chromatography, High Pressure Liquid / methods*
  • Diabetes Mellitus, Type 1 / blood
  • Diabetes Mellitus, Type 2 / blood
  • Freezing*
  • Glycated Hemoglobin / analysis*
  • Humans
  • Middle Aged

Substances

  • Glycated Hemoglobin A
  • hemoglobin A1c protein, human