Characterization of Al2O3 optically stimulated luminescence films for 2D dosimetry using a 6 MV photon beam

Phys Med Biol. 2016 Nov 7;61(21):7551-7570. doi: 10.1088/0031-9155/61/21/7551. Epub 2016 Oct 7.

Abstract

This work evaluates the dosimetric properties of newly developed optically stimulated luminescence (OSL) films, fabricated with either Al2O3:C or Al2O3:C,Mg, using a prototype laser scanning reader, a developed image reconstruction algorithm, and a 6 MV therapeutic photon beam. Packages containing OSL films (Al2O3:C and Al2O3:C,Mg) and a radiochromic film (Gafchromic EBT3) were irradiated using a 6 MV photon beam using different doses, field sizes, with and without wedge filter. Dependence on film orientation of the OSL system was also tested. Diode-array (MapCHECK) and ionization chamber measurements were performed for comparison. The OSLD film doses agreed with the MapCHECK and ionization chamber data within the experimental uncertainties (<2% at 1.5 Gy). The system background and minimum detectable dose (MDD) were <0.5 mGy, and the dose response was approximately linear from the MDD up to a few grays (the linearity correction was <10% up to ~2-4 Gy), with no saturation up to 30 Gy. The dose profiles agreed with those obtained using EBT3 films (analyzed using the triple channel method) in the high dose regions of the images. In the low dose regions, the dose profiles from the OSLD films were more reproducible than those from the EBT3 films. We also demonstrated that the OSL film data are independent on scan orientation and field size over the investigated range. The results demonstrate the potential of OSLD films for 2D dosimetry, particularly for the characterization of small fields, due to their wide dynamic range, linear response, resolution and dosimetric properties. The negligible background and potential simple calibration make these OSLD films suitable for remote audits. The characterization presented here may motivate further commercial development of a 2D dosimetry system based on the OSL from Al2O3:C or Al2O3:C,Mg.

MeSH terms

  • Algorithms*
  • Aluminum Oxide / chemistry*
  • Calibration
  • Dose-Response Relationship, Radiation
  • Film Dosimetry / instrumentation*
  • Film Dosimetry / methods
  • Luminescent Measurements / instrumentation*
  • Optics and Photonics*

Substances

  • Aluminum Oxide