Automated peak width measurements for targeted analysis of ion mobility unresolved species

Anal Chim Acta. 2016 Oct 19:941:49-60. doi: 10.1016/j.aca.2016.08.019. Epub 2016 Aug 15.

Abstract

Peak broadening in ion mobility (IM) is a relatively predictable process and abnormally broad peaks can be indicative of the presence of unresolved species. Here, we introduce a new ion mobility peak fitting (IM_FIT) software package for automated and systematic determination of traveling wave ion mobility (TWIM) unresolved species. To identify IM unresolved species, the IM_FIT software generates a trend line by plotting ions' mobility peak widths as a function of their arrival times. Utilizing user-defined thresholds, IM_FIT allows for automated and rapid detection of ions that deviate from the peak width trend line. To demonstrate the advantages of IM_FIT for automated detection of IM unresolved species, IM-mass spectrometry (IM-MS) data from a sample mixture containing polypropylene glycol and multiple peptides were analyzed. A total of 14 out of the 34 observed singly-charged IM peaks above 5% relative abundance (i.e., signal-to-noise ratios above ∼200) were tagged as potentially co-eluting ions by IM_FIT. Subsequently, the 14 IM peaks tagged as potentially unresolved (presumably, peaks corresponding to co-eluting compounds), were further analyzed by automated IM deconvolution (AIMD), liquid chromatography-IM-MS (LC-IM-MS), and/or ultra-high resolution mass spectrometry. Using the aforementioned techniques, more than 85% of the tagged IM peaks (12 out of 14) were confirmed to contain co-eluting ions. As an additional new finding, IM_FIT facilitated the discovery of an unexpected sequence-scrambled y-type fragment ion.

Keywords: Chemometrics; High-throughput; Ion mobility spectrometry (IMS); Mass spectrometry (MS); Mixture analysis; Software.

MeSH terms

  • Automation
  • Mass Spectrometry*
  • Software
  • Statistics as Topic / methods*
  • User-Computer Interface