Evaluation of a liquid culture system in the detection of mycobacteria at an antituberculosis institution in China; A retrospective study

J Int Med Res. 2016 Oct;44(5):1055-1060. doi: 10.1177/0300060516655243. Epub 2016 Sep 29.

Abstract

Objective A retrospective study comparing the diagnostic performance of the BACTEC™ MGIT™ 960 system (M960 system; BD Worldwide, Franklin Lakes, NJ, USA) with Löwenstein-Jensen (L-J) culture to detect mycobacteria in sputum specimens from patients with suspected pulmonary tuberculosis (TB). Methods Sputum samples were analysed for the presence of acid-fast bacilli (AFB). Samples were inoculated into the M960 system and L-J culture. Positive cultures were examined for the presence of AFB. Results The M960 method detected significantly more positive samples than L-J culture (818/1676 [48.8%] vs 692/1676 [41.3%]). Using L-J culture as reference, the sensitivity, specificity, positive predictive and negative predictive values of the M960 system were 91.0%, 76.1%, 77.0% and 92.2%, respectively. The time-to-detection of mycobacteria was 11.78 ± 5.16 days for M960 and 24.17 ± 8.73 days for L-J. Conclusions The M960 system had better diagnostic capability than L-J culture. Clinical value may be maximized by combining results from both methods.

Keywords: BACTEC MGIT 960; Liquid culture; Lowenstein–Jenson; mycobacteria; solid culture; tuberculosis.

Publication types

  • Comparative Study

MeSH terms

  • China
  • Culture Media
  • Humans
  • Mycobacterium tuberculosis / isolation & purification*
  • Retrospective Studies
  • Sputum / microbiology*
  • Tuberculosis, Pulmonary / diagnosis*
  • Tuberculosis, Pulmonary / microbiology