Routine evaluation of left ventricular function using CZT-SPECT, with low injected activities and limited recording times

J Nucl Cardiol. 2018 Feb;25(1):249-256. doi: 10.1007/s12350-016-0615-z. Epub 2016 Sep 27.

Abstract

Purpose: Gamma-cameras, with Cadmium-Zinc-Telluride (CZT) detectors, allow to perform myocardial perfusion imaging (MPI) with limited injected activities and recorded times. This study aimed at determining whether the routine assessment of left ventricular (LV) function with such limited counts protocols compares well with reference values from cardiac MRI.

Methods: The study included patients who have undergone cardiac MRI and an MPI routinely planned on a CZT camera with a low-dose protocol (120 MBq of Sestamibi for stress and 360 MBq at rest for 75 kg body weight), while targeting the recording of only 500 myocardial kcounts in order to limit the recording times (<10 minutes for stress, <4 minutes for rest). SPECT images were reconstructed with a method maintaining rather high spatial (8 mm) and temporal (16 frames/cycle) resolutions.

Results: Seventy-six patients were included, and mean effective dose was 3.5 ± 1.7 mSv for the total MPI protocol. Correlations between CZT-SPECT and MRI were good to excellent for ejection fraction (r 2 = 0.77), end-diastolic (r 2 = 0.88) and end-systolic (r 2 = 0.93) volumes, and the analysis of segmental contractility correlated well between the two techniques (kappa score = 0.72 ± 0.02).

Conclusion: LV function, assessed on a CZT camera with low injected activities and limited recording times, correlates well with the reference assessment from cardiac MRI.

Keywords: CZT camera; Myocardial perfusion imaging (MPI); Tc-99m Sestamibi; gated-SPECT; left ventricular function.

MeSH terms

  • Aged
  • Body Weight
  • Cadmium*
  • Female
  • Gamma Cameras
  • Heart / diagnostic imaging
  • Humans
  • Image Processing, Computer-Assisted / methods
  • Imaging, Three-Dimensional
  • Magnetic Resonance Imaging*
  • Male
  • Middle Aged
  • Myocardial Perfusion Imaging*
  • Radiation Dosage
  • Radiopharmaceuticals
  • Retrospective Studies
  • Systole
  • Technetium Tc 99m Sestamibi
  • Tellurium*
  • Time Factors
  • Tomography, Emission-Computed, Single-Photon*
  • Ventricular Function, Left*
  • Zinc*

Substances

  • CdZnTe
  • Radiopharmaceuticals
  • Cadmium
  • Technetium Tc 99m Sestamibi
  • Zinc
  • Tellurium