Randomized clinical trial comparing abluminal biodegradable polymer sirolimus-eluting stents with durable polymer sirolimus-eluting stents: Nine months angiographic and 5-year clinical outcomes

Medicine (Baltimore). 2016 Sep;95(38):e4820. doi: 10.1097/MD.0000000000004820.

Abstract

Background: The biodegradable polymer drug-eluting stents (DES) were developed to improve vascular healing. However, further data and longer-term follow-up are needed to confirm safety and efficacy of these stents. This randomized clinical trial aimed to compare safety and efficacy of 2 sirolimus-eluting stents (SES): Cordimax-a novel abluminal biodegradable polymer SES and Cypher Select-a durable polymer SES, at 9 months angiographic and 5-year clinical follow-up.

Methods: We randomized 402 patients with coronary artery disease to percutaneous coronary intervention with Cordimax (n = 202) or Cypher select (n = 200). Angiographic follow-up was performed at 9 months after the index procedure and clinical follow-up annually up to 5 years. The primary endpoint was angiographic in-stent late luminal loss (LLL). Secondary endpoints included angiographic restenosis rate, target vessel revascularization (TVR), and major adverse cardiac events (MACEs; defined as cardiac death, myocardial infarction, or TVR) at 5-year follow-up.

Results: Cordimax was noninferior to Cypher select for in-stent LLL (0.25 ± 0.47 vs 0.18 ± 0.49 mm; P = 0.587) and in-stent mean diameter stenosis (22.19 ± 12.21% vs 19.89 ± 10.79%; P = 0.064) at 9 months angiographic follow-up. The MACE rates were not different at 1 year (5.9% vs 4.0%, P = 0.376); however, MACE rates from 2 to 5 years were lower in the Cordimax group (6.8% vs 13.1%; P = 0.039).

Conclusion: Abluminal biodegradable polymer SES is noninferior to durable polymer SES at 9-month angiographic and 1-year clinical follow-up. However, MACE rates from 2 to 5 years were less in the abluminal biodegradable polymer group.

Publication types

  • Comparative Study
  • Randomized Controlled Trial

MeSH terms

  • Absorbable Implants
  • Aged
  • Coronary Angiography / statistics & numerical data*
  • Coronary Artery Disease / surgery*
  • Coronary Restenosis / epidemiology
  • Coronary Restenosis / etiology
  • Drug-Eluting Stents*
  • Durable Medical Equipment
  • Female
  • Follow-Up Studies
  • Humans
  • Immunosuppressive Agents / administration & dosage
  • Male
  • Middle Aged
  • Percutaneous Coronary Intervention / adverse effects
  • Percutaneous Coronary Intervention / instrumentation*
  • Polymers
  • Postoperative Complications / epidemiology
  • Postoperative Complications / etiology*
  • Prosthesis Design
  • Sirolimus / administration & dosage
  • Time Factors
  • Treatment Outcome

Substances

  • Immunosuppressive Agents
  • Polymers
  • Sirolimus