Calibration of a low background gas-flow proportional counter to estimate 234Th activity in coastal waters

Appl Radiat Isot. 2016 Dec:118:201-210. doi: 10.1016/j.apradiso.2016.09.016. Epub 2016 Sep 13.

Abstract

This paper relates the calibration of a low background gas-flow proportional counter. This calibration has been used to determine low activity of 234Th in coastal water samples. Two methods were used to prepare calibration samples: Evaporation and Electrodeposition. First method was rejected due to the lack of reproducibility because the different geometry adopted by the drops of tracer once dried on the disk. On the contrary, through the second method, similar efficiencies were obtained in all detectors with an average of 0.401±0.004. In this paper, the whole procedure to obtain 234Th activity in dissolution as well as in particulate matter has been detailed, and all the algorithms needed to calculate activities and efficiencies are shown. Finally, two experiments have been designed in order to validate the calibration of the beta counter and the method to determine 234Th in coastal waters with high concentration of particulate matter.

Keywords: Beta counting; Calibration; Methodology; Thorium-234.