Real-time magnetic resonance imaging guidance improves the diagnostic yield of endomyocardial biopsy

JACC Basic Transl Sci. 2016 Aug;1(5):376-383. doi: 10.1016/j.jacbts.2016.05.007.

Abstract

Background: Diagnostic yield of endomyocardial biopsy is low, particularly in disease that affects the myocardium in a non-uniform distribution. We hypothesized that real-time MRI guidance could improve the yield through targeted biopsy of focal myocardial pathology.

Methods: An animal model of focal myocardial pathology was created by infusing 3mL of fluorescent microspheres (NuFlow Hydrocoat, 15μm diameter, 5 million spheres/mL) followed by 2mL of 100% ethanol to a branch coronary artery. Animals were survived for minimum 14days, before undergoing MRI guided endomyocardial biopsy using a custom 6.5Fr active visualization MRI-conditional bioptome and X-ray guided biopsy using a commercial bioptome. Specimens were analyzed using a dissecting microscope under ultraviolet light to determine the proportion of 'on-target' specimens containing fluorescent microspheres.

Results: A total of 77 specimens were obtained using real-time MRI guidance and 87 using X-ray guidance, in five animals. Specimens obtained with the MRI-conditional bioptome were smaller compared with the commercial X-ray bioptome. Real-time MRI guidance significantly increased the diagnostic yield of endomyocardial biopsy (82% vs. 56% on-target biopsy specimens with real-time MRI vs. X-ray guidance, p<0.01).

Conclusions: Endomyocardial biopsy performed using real-time MRI guidance is feasible and significantly improves the diagnostic yield compared with X-ray fluoroscopy guidance.