Guidelines to electrode positioning for human and animal electrical impedance myography research

Sci Rep. 2016 Sep 2:6:32615. doi: 10.1038/srep32615.

Abstract

The positioning of electrodes in electrical impedance myography (EIM) is critical for accurately assessing disease progression and effectiveness of treatment. In human and animal trials for neuromuscular disorders, inconsistent electrode positioning adds errors to the muscle impedance. Despite its importance, how the reproducibility of resistance and reactance, the two parameters that define EIM, are affected by changes in electrode positioning remains unknown. In this paper, we present a novel approach founded on biophysical principles to study the reproducibility of resistance and reactance to electrode misplacements. The analytical framework presented allows the user to quantify a priori the effect on the muscle resistance and reactance using only one parameter: the uncertainty placing the electrodes. We also provide quantitative data on the precision needed to position the electrodes and the minimum muscle length needed to achieve a pre-specified EIM reproducibility. The results reported here are confirmed with finite element model simulations and measurements on five healthy subjects. Ultimately, our data can serve as normative values to enhance the reliability of EIM as a biomarker and facilitate comparability of future human and animal studies.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Adult
  • Animals
  • Biomedical Research*
  • Electric Impedance*
  • Electrodes
  • Humans
  • Movement
  • Muscle, Skeletal / physiology
  • Myography*
  • Practice Guidelines as Topic*