Clinical evaluation of the air-handling properties of contemporary oxygenators with integrated arterial filter

Perfusion. 2017 Mar;32(2):118-125. doi: 10.1177/0267659116664402. Epub 2016 Aug 20.

Abstract

Gaseous microemboli (GME) may originate from the extracorporeal circuit and enter the arterial circulation of the patient. GME are thought to contribute to cerebral deficit and to adverse outcome after cardiac surgery. The arterial filter is a specially designed component for removing both gaseous and solid microemboli. Integration of an arterial filter with an oxygenator is a contemporary concept, reducing both prime volume and foreign surface area. This study aims to determine the air-handling properties of four contemporary oxygenator devices with an integrated arterial filter. Two oxygenator devices, the Capiox FX25 and the Fusion, showed significant increased volume of GME reduction rates (95.03 ± 3.13% and 95.74 ± 2.69%, respectively) compared with both the Quadrox-IF (85.23 ± 5.84%) and the Inspire 6F M (84.41 ± 12.93%). Notably, both the Quadrox-IF and the Inspire 6F M as well as the Capiox FX 25 and the Fusion showed very similar characteristics in volume and number reduction rates and in detailed distribution properties. The Capiox FX25 and the Fusion devices showed significantly increased number and volume reduction rates compared with the Quadrox-IF and the Inspire 6F M devices. Despite the large differences in design of all four devices, our study results suggest that the oxygenator devices can be subdivided into two groups based on their fibre design, which results in screen filter (Quadrox-IF and Inspire 6F M) and depth filter (Capiox FX25 and Fusion) properties. Depth filter properties, as present in the Capiox FX25 and Fusion devices, reduced fractionation of air and may ameliorate GME removal.

Keywords: cardiopulmonary bypass; clinical evaluation; design; gaseous microemboli; oxygenators.

Publication types

  • Evaluation Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Arteries / surgery
  • Cardiopulmonary Bypass / adverse effects
  • Cardiopulmonary Bypass / instrumentation*
  • Embolic Protection Devices*
  • Embolism, Air / etiology
  • Embolism, Air / prevention & control*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Oxygenators* / adverse effects
  • Prospective Studies