In a previous work, phase-space data files (phsp) provided by the International Atomic Energy Agency (IAEA) were used to develop a hybrid virtual source model (VSM) for clinical photon beams. Very good agreement with dosimetric measurements performed on linear accelerators was obtained for field sizes up to 15×15cm(2). In the present work we extend the VSM to larger field sizes, for which phsp are not available. We incorporate a virtual flattening filter to our model, which can be determined from dose measurements for larger fields. In this way a fully functional VSM can be built, from publicly available IAEA's phsps and standard dose measurements, for fields of any size and tailored to a particular linac.
Keywords: IAEA; Large fields; Monte Carlo; Phase-space; Radiotherapy; Virtual source.
Copyright © 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.