Novel Triggered Nocturnal Blood Pressure Monitoring for Sleep Apnea Syndrome: Distribution and Reproducibility of Hypoxia-Triggered Nocturnal Blood Pressure Measurements

J Clin Hypertens (Greenwich). 2017 Jan;19(1):30-37. doi: 10.1111/jch.12878. Epub 2016 Jul 14.

Abstract

Obstructive sleep apnea (OSA) causes blood pressure (BP) surges during sleep, which may lead to increased sleep-onset cardiovascular events. The authors recently developed a triggered nocturnal BP monitoring system that initiates BP measurements when oxygen desaturation falls below a variable threshold. The distribution and reproducibility of hypoxia-triggered nocturnal BP parameters compared with those of fixed-interval nocturnal BP parameters for two consecutive nights in 147 OSA patients (mean age 59.4 years, 86.4% men) were evaluated. The mean and distribution (standard deviation [SD]) of the hypoxia-peak systolic BP (SBP) were significantly greater than that of the mean nocturnal SBP (mean±SD: 148.8±20.5 vs 123.4±14.2 mm Hg, P<.001). The repeatability coefficient (expressed as %MV) of hypoxia-peak SBP between night 1 and night 2 was comparable to that of mean nocturnal SBP (43% vs 32%). In conclusion, hypoxia-peak nocturnal BP was much higher than mean nocturnal BP, and it was as reproducible as mean nocturnal BP.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Blood Pressure
  • Blood Pressure Monitoring, Ambulatory / methods*
  • Female
  • Humans
  • Hypertension / physiopathology*
  • Hypoxia / physiopathology*
  • Male
  • Middle Aged
  • Reproducibility of Results
  • Sleep Apnea, Obstructive / physiopathology*