Application of Real-Time Tissue Elastography with a Low Frequency Convex Array Probe: A Noninvasive Approach to Differential Diagnosis of Liver Tumors

ISRN Hepatol. 2014 Apr 1:2014:378243. doi: 10.1155/2014/378243. eCollection 2014.

Abstract

To evaluate diagnostic performance of real-time tissue elastography (RTE) with a low frequency convex array probe for distinguishing benign from malignant hepatic tumors through trans-abdominal examination, elasticity images of 210 liver tumors were obtained by EUB-7500 (Hitachi Medical Systems and 3.5 MHz probe) and eventually 121 liver tumors were analyzed in the study. Elasticity images were classified into four types, from type a to d. Regarding type a or b as benign tumors and type c or d as malignant ones, sensitivity, specificity, and accuracy were calculated and the consistency between the findings of RTE and the pathohistological diagnosis was evaluated. The sensitivity, specificity, and accuracy were separately 97.2%, 88.0%, and 93.4% (P < 0.001). Moreover, there was a good consistency between the findings of RTE and the pathological diagnosis (kappa value 0.86). Among elasticity images of all the malignant tumors, the hepatocellular carcinomas (HCCs) mainly appeared in type c, and liver metastatic cancers in type d. Thus, RTE utilized as a novel noninvasive imaging examination method enables us to distinguish benign from malignant liver tumors. Moreover, it provides certain information for the differential diagnosis between HCCs and liver metastatic cancers.