[Impact of the Infrared Monitor Signal Pattern on Accuracy of Target Imaging in 4-dimensional Cone-beam Computed Tomography]

Nihon Hoshasen Gijutsu Gakkai Zasshi. 2016 Jun;72(6):469-79. doi: 10.6009/jjrt.2016_JSRT_72.6.469.
[Article in Japanese]

Abstract

To realize the high precision radiotherapy, localized radiation field of the moving target is very important, and visualization of a temporal location of the target can help to improve the accuracy of the target localization. However, conditions of the breathing and the patient's own motion differ from the situation of the treatment planning. Therefore, positions of the tumor are affected by these changes. In this study, we implemented a method to reconstruct target motions obtained with the 4D CBCT using the sorted projection data according to the phase and displacement of the extracorporeal infrared monitor signal, and evaluated the proposed method with a moving phantom. In this method, motion cycles and positions of the marker were sorted to reconstruct the image, and evaluated the image quality affected by changes in the cycle, phase, and positions of the marker. As a result, we realized the visualization of the moving target using the sorted projection data according to the infrared monitor signal. This method was based on the projection binning, in which the signal of the infrared monitor was surrogate of the tumor motion. Thus, further major efforts are needed to ensure the accuracy of the infrared monitor signal.

Publication types

  • English Abstract

MeSH terms

  • Cone-Beam Computed Tomography / standards*
  • Image Processing, Computer-Assisted*
  • Infrared Rays*
  • Phantoms, Imaging
  • Radiotherapy, Image-Guided / methods*