Corrosion of the Head-neck Junction After Total Hip Arthroplasty

J Am Acad Orthop Surg. 2016 Jun;24(6):349-56. doi: 10.5435/JAAOS-D-15-00111.

Abstract

Corrosion of the head-neck junction of implants used in total hip arthroplasty is a complex problem. Clinical severity appears to be multifactorial, and the predictive variables have yet to be consistently identified in the literature. Corrosion should be considered in the differential diagnosis of hip pain following total hip arthroplasty regardless of the type of bearing surface used. The most common presentation, pain followed by instability, is similar to complications associated with metal-on-metal articulations. The diagnosis of implant corrosion of the head-neck junction can be challenging; an infection workup should be performed along with analysis of serum metal ion levels and cross-sectional imaging. In the short term, a well-fixed stem may be retained, and the exchange of an isolated head with a ceramic femoral head seems to be a promising option for certain implants. Further research with longer follow-up is warranted, and high levels of evidence are needed to determine whether this approach is generalizable.

Publication types

  • Review

MeSH terms

  • Arthroplasty, Replacement, Hip / adverse effects*
  • Arthroplasty, Replacement, Hip / instrumentation
  • Arthroplasty, Replacement, Hip / methods
  • Corrosion
  • Femur Head
  • Hip Prosthesis*
  • Humans
  • Pain, Postoperative / diagnosis
  • Pain, Postoperative / etiology
  • Pain, Postoperative / surgery*
  • Prosthesis Design
  • Prosthesis Failure*
  • Reoperation / instrumentation
  • Reoperation / methods