A reliability study of the new sensors for movement analysis (SHARIF-HMIS)

J Bodyw Mov Ther. 2016 Apr;20(2):341-5. doi: 10.1016/j.jbmt.2015.10.004. Epub 2015 Oct 24.

Abstract

Aim: SHARIF-HMIS is a new inertial sensor designed for movement analysis. The aim of the present study was to assess the inter-tester and intra-tester reliability of some kinematic parameters in different lumbar motions making use of this sensor.

Materials and methods: 24 healthy persons and 28 patients with low back pain participated in the current reliability study. The test was performed in five different lumbar motions consisting of lumbar flexion in 0, 15, and 30° in the right and left directions. For measuring inter-tester reliability, all the tests were carried out twice on the same day separately by two physiotherapists. Intra-tester reliability was assessed by reproducing the tests after 3 days by the same physiotherapist.

Findings: The present study revealed satisfactory inter- and intra-tester reliability indices in different positions. ICCs for intra-tester reliability ranged from 0.65 to 0.98 and 0.59 to 0.81 for healthy and patient participants, respectively. Also, ICCs for inter-tester reliability ranged from 0.65 to 0.92 for the healthy and 0.65 to 0.87 for patient participants.

Conclusion: In general, it can be inferred from the results that measuring the kinematic parameters in lumbar movements using inertial sensors enjoys acceptable reliability.

Keywords: Inertial sensor; Inter-tester reliability; Intra-tester reliability; Kinematic parameters; SHARIF–HMIS.

MeSH terms

  • Adult
  • Biomechanical Phenomena
  • Humans
  • Low Back Pain / physiopathology*
  • Low Back Pain / rehabilitation*
  • Lumbar Vertebrae / physiopathology*
  • Male
  • Middle Aged
  • Movement / physiology*
  • Observer Variation
  • Physical Therapy Modalities / instrumentation*
  • Physical Therapy Modalities / standards
  • Range of Motion, Articular
  • Reproducibility of Results