The effect of dynamic hip motion on the micromotion of press-fit acetabular cups in six degrees of freedom

Med Eng Phys. 2016 Aug;38(8):717-24. doi: 10.1016/j.medengphy.2016.04.014.

Abstract

The hip joint is subjected to cyclic loading and motion during activities of daily living and this can induce micromotions at the bone-implant interface of cementless total hip replacements. Initial stability has been identified as a crucial factor to achieve osseointegration and long-term survival. Whilst fixation of femoral stems achieves good clinical results, the fixation of acetabular components remains a challenge. In vitro methods assessing cup stability keep the hip joint in a fixed position, overlooking the effect of hip motion. The effect of hip motion on cup micromotion using a hip motion simulator replicating hip flexion-extension and a six degrees of freedom measurement system was investigated. The results show an increase in cup micromotion under dynamic hip motion compared to Static Flexion. This highlights the need to incorporate hip motion and measure all degrees of freedom when assessing cup micromotion. In addition, comparison of two press-fit acetabular cups with different surface coatings suggested similar stability between the two cups. This new method provides a basis for a more representative protocol for future pre-clinical evaluation of different cup designs.

Keywords: Acetabular cups; Biomechanical study; Initial stability; Press-fit fixation; Total hip replacement.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Acetabulum / physiology*
  • Bone-Implant Interface
  • Hip Joint / physiology*
  • Hip Prosthesis*
  • Motion*
  • Movement*
  • Prosthesis Design