Technical Assessment of Connector-Assisted Nerve Repair

J Hand Surg Am. 2016 Jul;41(7):760-6. doi: 10.1016/j.jhsa.2016.04.015. Epub 2016 May 14.

Abstract

Purpose: Clinical outcomes of nerve repair have not substantially improved over the last several decades. Although clearly a multifactorial problem, basic principles including proper fascicular alignment are not always realized. The use of short conduits as connectors may enhance nerve alignment by entubulating and directing the approximation of nerve ends.

Methods: Ten hand surgeons (5 experienced and 5 inexperienced) performed a series of in vitro human cadaver nerve repairs. Three small-diameter (2 mm), 3 medium-diameter (3-4 mm), and 3 large-diameter (5-6 mm) nerves were repaired (under 10× magnification) utilizing each of 3 techniques: suture-only, connector-only (sutures placed through the ends of the connector), and connector-assisted (alignment sutures at the nerve interface plus connector). Three judges (blinded to who performed the repairs) assessed each repair for fascicular alignment based on predetermined qualitative scales.

Results: Across all surgeons, 23 of 30 connector-assisted repairs were judged good or excellent versus 18 of 30 of the suture-only and 13 of 30 of the connector-only repairs. Experienced surgeons in general did better repairs and in particular were more likely to obtain superior alignment for conduit-only repairs (73.3% vs 13.3% good or excellent) and suture-only repairs (73.3% vs 46.7% good or excellent) and were not statistically different for connector-assisted repairs (86.7% vs 66.7% good or excellent) compared with inexperienced surgeons.

Conclusions: In a cadaver nerve model, there was no significant difference in the technical alignment of conduit-assisted repairs between experienced and inexperienced surgeons whereas inexperienced surgeons were more likely to achieve inadequate alignment with suture-only or conduit-only repairs.

Clinical relevance: Connector-assisted repairs combining suture-approximation and entubulation may improve the technical alignment of nerve repairs performed, especially by less-experienced surgeons.

Keywords: connector-assisted; microsuture; nerve connector; nerve repair.

MeSH terms

  • Cadaver
  • Clinical Competence
  • Humans
  • Microsurgery / instrumentation*
  • Neurosurgical Procedures / instrumentation*
  • Peripheral Nerve Injuries / surgery*
  • Prostheses and Implants*
  • Sutures*