Hematocrit (HCT) measurements are important clinical diagnostic variables that help physicians diagnose and treat various medical conditions, ailments, and diseases. In this work, we present the HCT Disc, a centrifugal microdevice fabricated by a Print, Cut and Laminate (PCL) method to generate a 12-sample HCT device from materials costing <0.5 USD (polyester and toner or PeT). Following introduction from a drop of blood (finger stick), whole blood metering and cell sedimentation are controlled by centrifugal force, only requiring a CD player motor as external hardware and, ultimately, a cell phone for detection. The sedimented volume from patient blood in the HCT Disc was analyzed using a conventional scanner/custom algorithm for analysis of the image to determine a hematocrit value, and these were compared to values generated in a clinical laboratory, which correlated well. To enhance portability and assure simplicity of the HCT measurement, values from image analysis by a cell phone using a custom application was compared to the scanner. Fifteen samples were analyzed with cell phone image analysis system and were found to be within 4% of the HCT values determined in the clinical lab. We demonstrate the feasibility of the PeT device for HCT measurement, and highlight its uniquely low cost (<0.5 USD), speed (sample-to-answer <8 min), multiplexability (12 samples), low volume whole blood requirement (<3 μL), rotation speeds (<4000 rpm) needed for effective measurement as well as the direct finger-to-chip sample loading capability.
Keywords: Cell phone detection; Centrifugal microfluidics; Hematocrit; Polyester-toner.