EDTA interference in electrochemiluminescence ACTH assay

Ann Clin Biochem. 2016 Nov;53(6):699-701. doi: 10.1177/0004563216636898. Epub 2016 Sep 28.

Abstract

Background As plasma is the recommended sample type for Roche adrenocorticotropic hormone (ACTH) assay, we evaluated the effect of EDTA concentration on Cobas ACTH assay. Methods Samples containing twofold and fourfold higher concentrations of EDTA were prepared by adding plasma to empty K2EDTA tubes and by making under-filled EDTA tubes. All measurements were performed with four replicates. Results Increased EDTA concentration resulted in a significant decrease in ACTH concentration. Fifty-per cent-filled EDTA tube showed 19% decrease in ACTH concentration and 25% filled EDTA tube showed 50% decrease in ACTH concentration. Conclusion We recommend that inadequately filled EDTA specimens should be rejected when using Cobas ACTH assay.

Keywords: ACTH; EDTA; plasma.