Amperometric biosensor based on Laccase immobilized onto a screen-printed electrode by Matrix Assisted Pulsed Laser Evaporation

Talanta. 2016 Jul 1:154:438-45. doi: 10.1016/j.talanta.2016.03.072. Epub 2016 Mar 24.

Abstract

A Laccase-based biosensor for the determination of phenolic compounds was developed by using Matrix Assisted Pulsed Laser Evaporation as an innovative enzyme immobilization technique. and the deriving biosensor was characterized and applied for the first time. Laccase was immobilized onto different substrates including screen printed carbon electrodes and spectroscopic, morphologic and electrochemical characterizations were carried out. A linear range from 1 to 60μM was achieved working at 5.5pH and -0.2V detection potential vs Ag pseudoreference. The limits of detection and quantification were found to be 1 and 5μM, respectively. A good fabrication reproducibility, stability of response and selectivity toward interferents were also found The potential of the developed biosensor was tested in the determination of total polyphenol content in real matrices (tea infusion, ethanolic extract from Muscari comosum bulbs and aqueous solution of a food supplement from black radish root and artichoke leaves) and the results were compared with those obtained by using the Folin-Ciocalteu method.

Keywords: Biosensor; Laccase; Matrix Assisted Pulsed Laser Evaporation; Polyphenols assessment; Screen Printed Electrode.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biosensing Techniques*
  • Electrodes
  • Enzymes, Immobilized
  • Laccase
  • Reproducibility of Results
  • Volatilization

Substances

  • Enzymes, Immobilized
  • Laccase