Visual Evoked Potential Using Head-Mounted Display Versus Cathode Ray Tube: A Pilot Study

Ann Rehabil Med. 2016 Apr;40(2):334-40. doi: 10.5535/arm.2016.40.2.334. Epub 2016 Apr 25.

Abstract

Objective: To present a new stimulation method based on the use of a head-mounted display (HMD) during pattern reversal visual evoked potential (PR-VEP) testing and to compare variables of HMD to those of conventional cathode ray tube (CRT).

Methods: Twenty-three normal subjects without visual problems were recruited. PR-VEPs were generated using CRT or HMD stimuli. VEP outcome measures included latencies (N75, P100, and N145) and peak-to-peak amplitudes (N75-P100 and P100-N145). Subjective discomfort associated with HMD was determined using a self-administered questionnaire.

Results: PR-VEPs generated by HMD stimuli showed typical triphasic waveforms, the components of which were found to be correlated with those obtained using conventional CRT stimuli. Self-administered discomfort questionnaires revealed that HMD was more comfortable in some aspects. It allowed subjects to concentrate better than CRT.

Conclusion: The described HMD stimulation can be used as an alternative to the standard CRT stimulation for PR-VEPs. PR-VEP testing using HMD has potential applications in clinical practice and visual system research because HMD can be used on a wider range of subjects compared to CRT.

Keywords: Cathode ray tube; Head-mounted display; Visual evoked potentials.