Assessment of pulse arrival time for arterial stiffness monitoring on body composition scales

Comput Biol Med. 2017 Jun 1:85:135-142. doi: 10.1016/j.compbiomed.2016.04.012. Epub 2016 Apr 22.

Abstract

This study presents a system that aims to estimate changes in arterial health status in an unobtrusive way. It might be especially useful in long-term self-monitoring of cardiovascular performance for successful treatment and empowerment of patients. This system applies the electrocardiographic and impedance plethysmographic signals acquired using modified body composition scales for the calculation of pulse arrival time, which is directly related to arterial stiffness. The proposed device was tested in a cohort of 14 subjects. The modified scales were compared to the commercial PulsePen tonometer and the results showed significant relationship between these different devices (rs=0.93, p<0.01). The system also showed the ability to track small pulse arrival time variations induced by paced respiration. These findings suggest that scales evaluating parameters of cardiovascular function have potential to become a convenient device for self-monitoring of arterial stiffness.

Keywords: Impedance plethysmogram; Multi-sensing device; Personal health technologies; Slope sum function; Unobtrusive long-term health monitoring.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Body Composition / physiology
  • Electrocardiography / methods*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Monitoring, Physiologic / methods
  • Plethysmography, Impedance / methods*
  • Pulse Wave Analysis / methods*
  • Reproducibility of Results
  • Signal Processing, Computer-Assisted*
  • Tibial Arteries / physiology
  • Vascular Stiffness / physiology*
  • Young Adult