A flexible pressure sensor could correctly measure the depth of chest compression on a mattress

Am J Emerg Med. 2016 May;34(5):899-902. doi: 10.1016/j.ajem.2016.02.052. Epub 2016 Feb 26.

Abstract

Background: Feedback devices are used to improve the quality of chest compression (CC). However, reports have noted that accelerometers substantially overestimate depth when cardiopulmonary resuscitation (CPR) is performed on a soft surface. Here, we determined whether a flexible pressure sensor could correctly evaluate the depth CC performed on a mannequin placed on a mattress.

Methods: Chest compression was performed 100 times/min by a compression machine on the floor or a mattress, and the depth of CC was monitored using a flexible pressure sensor (Shinnosukekun) and CPRmeter(™). The depth of machine-performed CC was consistently 5cm. We compared data from the feedback sensor with the true depth of CC using dual real-time auto feedback system that incorporated an infrared camera (CPR evolution(™)).

Results: On the floor, the true depth of CC was 5.0±0.0cm (n=100), or identical to the depth of CC performed by the machine. The Shinnosukekun(™) measured a mean (±SD) CC depth of 5.0±0.1cm (n=100), and the CPRmeter(™) measured a depth of 5.0±0.2cm (n=100). On the mattress, the true depth of CC was 4.4±0.0cm (n=100). The Shinnosukekun(™) measured a mean CC depth of 4.4±0.0cm (n=100), and the CPRmeter(™) measured a depth of 4.7±0.1cm (n=100). The data of CPRmeter(™) were overestimated (P<.0001 between the true depth and the CPRmeter(™)-measured depth).

Conclusion: The Shinnosukekun(™) could correctly measure the depth of CC on a mattress. According to our present results, the flexible pressure sensor could be a useful feedback system for CC performed on a soft surface.

Publication types

  • Evaluation Study

MeSH terms

  • Beds*
  • Equipment Design
  • Heart Massage / methods*
  • Humans
  • Manikins
  • Monitoring, Physiologic / instrumentation*
  • Pressure*